Optical Pyrometer
Consignes
Build the optical path with the filament in the image plane
Build the optical path with the filament in the image plane
The filament and the target must appear at the same focus, or you cannot compare them.
- Fit an objective lens at one end of a blackened tube, focusing the target onto a plane inside.
- Mount a small tungsten filament lamp at exactly that plane, so the filament sits IN the focused image.
- Add an eyepiece behind it, focused on the filament.
- Look through and confirm you see the filament sharply, superimposed on a sharp image of whatever the objective is pointed at.
Both must be sharp together. If the filament is in focus and the target is not, you are comparing a bright wire against a blur and the match point becomes vague. Getting the two into one plane is the whole optical design, and it is why the instrument has a fixed working distance range.
Blacken the tube's interior thoroughly. Stray light reaching the eyepiece raises the apparent background brightness and makes the filament vanish too early, reading low.Matériaux pour cette étape :
Acrylic Rod1 pièce
Brass Tube 1/2" OD1 pièce
LED - Basic 5mm1 pièceOutils nécessaires :
Cordless Drill/Driver (20V)
Drill Bit Set (29-Piece, HSS)
File Set
Digital Caliper 6-Inch
Hacksaw Frame with Blades (10-Pack)Wire the filament current control and read it
Wire the filament current control and read it
Filament current is the measurement. Everything else is optics.
- Power the filament from an adjustable supply through a fine potentiometer.
- Put an ammeter in series so the current is read precisely.
- Turn the knob and watch the filament go from black through dull red to bright yellow.
- Note the current at several visible colours.
The dial reads current and is LABELLED in temperature. There is no thermal sensor anywhere in the instrument — the scale is a calibration, established once by pointing the pyrometer at sources of known temperature. That is the same arrangement as the airspeed indicator in the pitot blueprint: a pressure gauge with a velocity scale printed on it.
Use a fine potentiometer and a stable supply. The match point is judged by eye and the eye is very sensitive to brightness differences, so a coarse control that jumps past the match makes the instrument far less precise than it could be.Matériaux pour cette étape :
1/4W Resistor Kit (600pcs, 30 Values)1 kit
Adjustable Bench Power Supply (30V/5A)1 pièce
Copper Wire (20 Gauge)1 rouleauOutils nécessaires :
Digital Multimeter (Lab Grade)
Analog Multimeter
Digital Caliper 6-InchAdd the red filter — and understand why it is essential
Add the red filter — and understand why it is essential
Comparing brightness across different colours is unreliable. Remove the colour.
- Fit a deep red filter between the filament and the eyepiece.
- Look again: both the filament and the target now appear in the same narrow band of red.
- Compare the ease of judging the match with and without the filter.
Without the filter you are comparing an orange filament against a yellow furnace and guessing. With it, both are red and only their BRIGHTNESS differs, which the eye judges far better than it judges colour. There is a second reason too: the physics relating brightness to temperature depends on wavelength, so the calibration is only valid if the instrument always looks at the same wavelength. The filter enforces that.
This is a general instrumentation move — remove every variable except the one you are measuring. The filter throws away most of the light and makes the instrument better.Matériaux pour cette étape :
Acrylic Sheet (Clear, 1/4 inch, 12x12)1 feuille
Food Colouring1 bouteilleOutils nécessaires :
File Set
Digital Caliper 6-Inch
Coping SawCalibrate, then find the emissivity error
Calibrate, then find the emissivity error
The instrument assumes the target radiates like a perfect black body. Real surfaces do not.
- Point it at a kiln with a known thermocouple reading and set your scale against it.
- Now heat two objects to the same temperature: one dull black oxidised steel, one bright polished steel.
- Read both with the pyrometer.
- Compare against a thermocouple touching each.
Matériaux pour cette étape :
Aluminum Flat Bar (1x1/4 inch, 36-inch)1 pièce
Graph Paper1 padOutils nécessaires :
Infrared Thermometer
Digital Multimeter (Lab Grade)
Kiln
Digital Caliper 6-InchThe null method, and history
The null method, and history
Ludwig Holborn and Ferdinand Kurlbaum published the disappearing-filament pyrometer in 1901 at the Physikalisch-Technische Reichsanstalt, the German national standards laboratory. Earlier optical pyrometers compared brightness against a standard lamp side by side; putting the filament INTO the image so it visually disappears is what made the judgement easy and the instrument practical.
Null methods are the most accurate kind of measurement, and this is a fine example. You are not asking the observer to judge how bright something is — a hopeless task — but only whether two things are the SAME brightness, which the eye does extremely well. The Wheatstone bridge in the resistance thermometer works identically: it does not measure resistance, it detects when a difference is zero. Whenever you can turn a measurement into the detection of a null, do it.
Where it sits in the chain: Wedgwood's pyrometer read a permanent effect after the fact. The thermocouple gave a live signal but had to be inserted into the process and eventually degrades there. The optical pyrometer removes contact entirely, so it can read a steel ladle, a glass furnace or a filament in a vacuum tube — anything visible and hot enough to glow.
Its honest limits: it only works above roughly 700 °C, where objects glow visibly; it depends on emissivity, which is often unknown; smoke, dust and dirty windows between instrument and target all read low; and the classic version depends on an observer's eye, so two operators may differ slightly. Modern radiation pyrometers replace the eye with a detector but inherit the emissivity problem completely.
Matériaux
10- 1 pièceEspace réservé
- 1 pièceEspace réservé
- 1 pièce€1.00
- Espace réservé
- Espace réservé
- 1 rouleauEspace réservé
- Espace réservé
- 1 bouteilleEspace réservé
- Espace réservé
- 1 padEspace réservé
Outils requis
10- Espace réservé
- Espace réservé
- Espace réservé
- Espace réservé
- Espace réservé
- Espace réservé
- Espace réservé
- Espace réservé
Blueprints liés
Ces blueprints partagent des connaissances — techniques, matériaux ou principes
CC0 Domaine public
Ce blueprint est publié sous CC0. Vous êtes libre de copier, modifier, distribuer et utiliser ce travail pour tout usage, sans demander la permission.
Soutenez le Maker en achetant des produits via son Blueprint où il perçoit une Commission Maker définie par les Vendeurs, ou créez une nouvelle itération de ce Blueprint et incluez-le comme connexion dans votre propre Blueprint pour partager les revenus.