Building a Gas Chromatograph
Every separation so far has used a liquid to carry the sample. In 1952 Archer Martin and Anthony James — Martin again, eight years after paper chromatography — used a GAS instead, and the whole business got faster by a factor of a hundred.
A gas is thin. Molecules diffuse through it quickly, so they find the stationary phase, partition into it and come back out again thousands of times a second rather than thousands of times an hour. That speed is why a gas chromatograph resolves a mixture in minutes where a column takes an afternoon, and it is why the method now sits in every forensic, environmental and flavour laboratory on earth.
The catch is that the sample must survive being vaporised, so this works for what boils without decomposing: solvents, fuels, essential oils. You will build a working instrument from copper tube, an aquarium pump and a thermistor, and separate the components of a lavender oil you distilled yourself.
고급
2 days
안내
1
1
Understand the three parts before you build any of them
Understand the three parts before you build any of them
A gas chromatograph is an injector, a column and a detector, with a carrier gas flowing through all three and heat holding the column at a steady temperature.
The carrier gas does nothing chemically — it only pushes. The column is a long tube whose inside is coated with a high-boiling liquid: THAT liquid is the stationary phase and it does the separating. The detector reports when something other than pure carrier gas comes out the far end. Everything expensive about a commercial instrument is refinement of those three; nothing about them is beyond a workshop.
2
2
Make a packed column
Make a packed column
Crush firebrick, sieve it to a narrow size range, and soak the granules in silicone grease thinned with acetone so each one carries a thin film. Spread it out until the acetone has gone completely, then pack it into 2 metres of 4 mm copper tube, tapping constantly, and coil the tube.
The brick is only a support with a huge surface area; the grease film is the actual stationary phase. Sieving matters more than it sounds — mixed grain sizes pack unevenly, gas finds a fast channel, and every peak comes out broad. Plug both ends with a little glass wool so the packing cannot migrate into your pump or your detector.
이 단계의 재료:
동관2 metre
내화 벽돌1 개
아세톤100 ml
글라스울 단열재1 팩필요한 도구:
디지털 저울
니트릴 장갑3
3
Heat the column evenly, and measure the temperature you actually have
Heat the column evenly, and measure the temperature you actually have
Wind nichrome wire evenly along the coiled column, insulate it, and drive it from a variable supply. Bond a thermocouple to the tube itself, not to the air near it.
Retention time depends steeply on temperature — roughly halving for every 20 to 30 °C you add — so a column that is hotter at one end than the other smears every peak, and a column whose temperature drifts between runs gives retention times you cannot compare. Even winding and a real measurement of the metal are what make the instrument repeatable rather than merely working.
이 단계의 재료:
니크롬선5 metre필요한 도구:
직류 전원 장치
MAX6675 모듈을 갖춘 K형 열전대
멀티미터4
4
Build a thermal conductivity detector
Build a thermal conductivity detector
Mount two matched thermistors in small chambers, one in the gas leaving the column and one in carrier gas that has bypassed it, and wire them as two arms of a bridge. Feed the bridge a steady low voltage and read the imbalance.
Each thermistor is self-heated by the current through it and cooled by the gas around it. Pure carrier gas cools both equally and the bridge sits balanced. The moment a different molecule emerges, the gas around one thermistor conducts heat differently, that bead changes temperature, its resistance changes and the bridge tips. This is a universal detector — it sees anything that is not the carrier — and it was the detector on the first commercial instruments.
이 단계의 재료:
NTC 서미스터 키트1 키트필요한 도구:
직류 전원 장치
멀티미터
악어클립 시험선5
5
Set a steady carrier flow — steady matters more than fast
Set a steady carrier flow — steady matters more than fast
Run an aquarium pump through a ballast vessel and a needle valve into the injector end, and measure the flow at the outlet by timing a soap bubble up a graduated tube.
The ballast is what turns a pump's pulsing into a smooth stream, and a pulsing carrier makes the detector baseline oscillate so hard that small peaks disappear into it. Air works as a carrier for this build; it is not ideal, because oxygen slowly degrades the stationary phase at temperature and its thermal conductivity is close to many analytes, which costs sensitivity. Nitrogen is better if you have it, helium better still.
필요한 도구:
공기 펌프
스톱워치
붕규산 비커6
6
Inject fast, inject tiny
Inject fast, inject tiny
Fit a septum at the column inlet, heat that inlet above the boiling point of your sample, and inject a fraction of a microlitre with a fine syringe in one quick motion.
Everything about injection is about making the sample enter as a narrow plug. A slow injection feeds the column over several seconds and every peak is that many seconds wide before separation even begins. Too much sample overloads the stationary phase and gives leaning, tailing peaks whose retention times shift with how much you put in — which quietly destroys the identification the whole instrument exists to provide.
필요한 도구:
실험실용 주사기7
7
Run a lavender oil, and read the chromatogram
Run a lavender oil, and read the chromatogram
Inject a dilute solution of lavender essential oil and log the detector voltage against time. You should see a solvent peak early, then a series of peaks over several minutes: linalool and linalyl acetate are the two largest components of real lavender oil.
Retention time identifies; peak area quantifies. Run a known pure compound to find its retention time, then look for that time in your unknown. This is exactly how a laboratory tells adulterated essential oil from genuine — a cheap oil cut with synthetic linalool shows the right peak in the wrong proportion to its neighbours.
이 단계의 재료:
에탄올 - 실험실 등급, 95 %50 ml필요한 도구:
실험실용 주사기
멀티미터
스톱워치8
8
Van Deemter: why there is a best flow rate
Van Deemter: why there is a best flow rate
Jupyter 노트북 불러오는 중…
필요한 도구:
스톱워치재료
8- 2 metre플레이스홀더
- 1 개플레이스홀더
- 50 g플레이스홀더
- 100 ml플레이스홀더
- 1 팩플레이스홀더
- 5 metre플레이스홀더
- 1 키트플레이스홀더
- 50 ml플레이스홀더
관련 블루프린트
이 블루프린트들은 지식을 공유합니다 — 기술, 재료 또는 원리
CC0 퍼블릭 도메인
이 블루프린트는 CC0로 공개되었습니다. 어떤 목적으로든 자유롭게 복사, 수정, 배포 및 사용할 수 있습니다.
제품 구매를 통해 메이커를 지원하세요. 판매자가 설정한 메이커 커미션 을 받거나, 이 블루프린트의 새로운 반복을 만들어 연결로 포함시킬 수 있습니다.