Bouwers Rotating-Anode X-Ray Tube
Coolidge's tube of 1913 solved the electrical problem: a heated filament sets the current, so the operator finally controls the beam instead of negotiating with a bag of gas. What it handed over was a **thermal** problem, and a brutal one.
Only about one per cent of the power you put into an X-ray tube leaves as X-rays. The rest becomes heat, delivered into a focal spot a couple of millimetres across. At 100 kV and 400 mA that is roughly 40 kilowatts arriving in an area smaller than a grain of rice - hundreds of times the power density of a hotplate, into tungsten. The spot does not gradually warm. It pits, and then it melts, and a pitted anode makes a blurred and asymmetric beam for the rest of the tube's life.
Albert Bouwers' answer at Philips was to stop letting the beam hit the same place twice. **US 1,933,005** "X-ray tube", filed 13. March 1930 and granted 31. October 1933: put the target on a disc and spin it, so the beam traces a ring of fresh metal instead of a point. The patent says why plainly - "since the bombarded surface of the anticathode is materially larger than in tubes of the type having a concentrated focal spot which is formed on a stationary anticathode, such tubes present the advantage that a much higher load may be applied". It sold as the Rotalix from 1929.
**The clever part is how it is driven.** You cannot run a shaft through the wall of a vacuum tube and still have a vacuum. So the anode is mounted on a rotor sealed INSIDE the envelope, and the patent claims "a metallic stator carrying magnetic coils for imparting rotative motion to said anode and arranged without the tube" - an induction motor whose stator is outside the glass and whose rotor is inside it, coupled by nothing but a magnetic field. Nothing passes through the wall at all.
**That is what you will build.** Not an X-ray tube - this blueprint asks nobody to make ionising radiation - but the actual mechanism Bouwers patented: a stator outside a sealed glass jar spinning a rotor inside it, with no connection between them. It is the same motor, at the same working principle, and you will measure its speed, its slip and its run-up time.
درمیانہ
3 hours
خطرناک مواد
اس بلو پرنٹ میں خطرناک طریقہ کار شامل ہے۔ قدم بہ قدم ہدایات دیکھنے کے لیے لاگ ان کریں اور اپنی اکاؤنٹ سیٹنگز میں خطرناک مواد فعال کریں۔
متعلقہ بلیو پرنٹ
یہ بلیو پرنٹ علم بانٹتے ہیں — تکنیک، مواد یا اصول
CC0 پبلک ڈومین
یہ بلیو پرنٹ CC0 کے تحت جاری کیا گیا ہے۔ آپ اجازت لیے بغیر اس کام کو نقل، ترمیم، تقسیم اور کسی بھی مقصد کے لیے استعمال کرنے کے لیے آزاد ہیں۔
میکر کی حمایت کریں ان کے بلیو پرنٹ کے ذریعے پروڈکٹس خرید کر جہاں وہ میکر کمیشن وینڈرز کی طرف سے مقرر، کماتے ہیں، یا اس بلیو پرنٹ کی نئی تکرار بنائیں اور آمدنی شیئر کرنے کے لیے اسے اپنے بلیو پرنٹ میں کنکشن کے طور پر شامل کریں۔