Optical Pyrometer
Penny

작성자

Penny

26. 8월 2026DK
67
0
0
0
0

Optical Pyrometer

Every temperature instrument so far has to touch the thing it measures, and above about 1500 °C nothing survives contact for long. The optical pyrometer never touches anything. Look through it at a furnace and you see a fine electric filament superimposed on the glowing interior; turn a knob to change the filament's current until it vanishes against the background, and read the temperature off the dial. The filament disappears when it is radiating at exactly the same brightness as the target, so the measurement is a null comparison — you are not measuring light at all, you are matching two brightnesses and reading a current you can calibrate. Holborn and Kurlbaum built the disappearing-filament instrument in 1901, and it made molten steel measurable from across the room.
고급
4 hours 30 minutes

안내

1

Build the optical path with the filament in the image plane

The filament and the target must appear at the same focus, or you cannot compare them.

  1. Fit an objective lens at one end of a blackened tube, focusing the target onto a plane inside.
  2. Mount a small tungsten filament lamp at exactly that plane, so the filament sits IN the focused image.
  3. Add an eyepiece behind it, focused on the filament.
  4. Look through and confirm you see the filament sharply, superimposed on a sharp image of whatever the objective is pointed at.

Both must be sharp together. If the filament is in focus and the target is not, you are comparing a bright wire against a blur and the match point becomes vague. Getting the two into one plane is the whole optical design, and it is why the instrument has a fixed working distance range.

Blacken the tube's interior thoroughly. Stray light reaching the eyepiece raises the apparent background brightness and makes the filament vanish too early, reading low.

이 단계의 재료:

아크릴 봉아크릴 봉1 개
황동관 바깥지름 1/2인치황동관 바깥지름 1/2인치1 개
LED - Basic 5mmLED - Basic 5mm1 개

필요한 도구:

무선 드릴무선 드릴
드릴 비트 세트드릴 비트 세트
줄 세트줄 세트
디지털 캘리퍼스 6인치디지털 캘리퍼스 6인치
쇠톱 프레임과 톱날쇠톱 프레임과 톱날
2

Wire the filament current control and read it

Filament current is the measurement. Everything else is optics.

  1. Power the filament from an adjustable supply through a fine potentiometer.
  2. Put an ammeter in series so the current is read precisely.
  3. Turn the knob and watch the filament go from black through dull red to bright yellow.
  4. Note the current at several visible colours.

The dial reads current and is LABELLED in temperature. There is no thermal sensor anywhere in the instrument — the scale is a calibration, established once by pointing the pyrometer at sources of known temperature. That is the same arrangement as the airspeed indicator in the pitot blueprint: a pressure gauge with a velocity scale printed on it.

Use a fine potentiometer and a stable supply. The match point is judged by eye and the eye is very sensitive to brightness differences, so a coarse control that jumps past the match makes the instrument far less precise than it could be.

이 단계의 재료:

1/4 W 저항 키트1/4 W 저항 키트1 키트
가변 실험용 전원 장치가변 실험용 전원 장치1 개
구리선구리선1 롤

필요한 도구:

실험실급 디지털 멀티미터실험실급 디지털 멀티미터
아날로그 멀티미터아날로그 멀티미터
디지털 캘리퍼스 6인치디지털 캘리퍼스 6인치
3

Add the red filter — and understand why it is essential

Comparing brightness across different colours is unreliable. Remove the colour.

  1. Fit a deep red filter between the filament and the eyepiece.
  2. Look again: both the filament and the target now appear in the same narrow band of red.
  3. Compare the ease of judging the match with and without the filter.

Without the filter you are comparing an orange filament against a yellow furnace and guessing. With it, both are red and only their BRIGHTNESS differs, which the eye judges far better than it judges colour. There is a second reason too: the physics relating brightness to temperature depends on wavelength, so the calibration is only valid if the instrument always looks at the same wavelength. The filter enforces that.

This is a general instrumentation move — remove every variable except the one you are measuring. The filter throws away most of the light and makes the instrument better.

이 단계의 재료:

아크릴 판아크릴 판1 장
식용 색소식용 색소1 병

필요한 도구:

줄 세트줄 세트
디지털 캘리퍼스 6인치디지털 캘리퍼스 6인치
실톱(코핑 소)실톱(코핑 소)
4

Calibrate, then find the emissivity error

The instrument assumes the target radiates like a perfect black body. Real surfaces do not.

  1. Point it at a kiln with a known thermocouple reading and set your scale against it.
  2. Now heat two objects to the same temperature: one dull black oxidised steel, one bright polished steel.
  3. Read both with the pyrometer.
  4. Compare against a thermocouple touching each.
The polished surface reads far too LOW, because it emits less radiation at the same temperature — its emissivity is lower. This is the fundamental limitation of every non-contact temperature measurement, including modern infrared thermometers. The standard workaround is to sight into a deep cavity or a furnace port, because a small opening into a large enclosure behaves almost exactly like a black body regardless of what the walls are made of.

이 단계의 재료:

알루미늄 평각봉알루미늄 평각봉1 개
모눈종이모눈종이1 pad

필요한 도구:

적외선 온도계적외선 온도계
실험실급 디지털 멀티미터실험실급 디지털 멀티미터
가마(킬른)가마(킬른)
디지털 캘리퍼스 6인치디지털 캘리퍼스 6인치
5

The null method, and history

Ludwig Holborn and Ferdinand Kurlbaum published the disappearing-filament pyrometer in 1901 at the Physikalisch-Technische Reichsanstalt, the German national standards laboratory. Earlier optical pyrometers compared brightness against a standard lamp side by side; putting the filament INTO the image so it visually disappears is what made the judgement easy and the instrument practical.

Null methods are the most accurate kind of measurement, and this is a fine example. You are not asking the observer to judge how bright something is — a hopeless task — but only whether two things are the SAME brightness, which the eye does extremely well. The Wheatstone bridge in the resistance thermometer works identically: it does not measure resistance, it detects when a difference is zero. Whenever you can turn a measurement into the detection of a null, do it.

Where it sits in the chain: Wedgwood's pyrometer read a permanent effect after the fact. The thermocouple gave a live signal but had to be inserted into the process and eventually degrades there. The optical pyrometer removes contact entirely, so it can read a steel ladle, a glass furnace or a filament in a vacuum tube — anything visible and hot enough to glow.

Its honest limits: it only works above roughly 700 °C, where objects glow visibly; it depends on emissivity, which is often unknown; smoke, dust and dirty windows between instrument and target all read low; and the classic version depends on an observer's eye, so two operators may differ slightly. Modern radiation pyrometers replace the eye with a detector but inherit the emissivity problem completely.

재료

10

필요 도구

10
예상 총액
만든 사람이 산 것. 가격이 없는 재료는 구매하는 곳에서 구하면 됩니다.
₩10

관련 블루프린트

이 블루프린트들은 지식을 공유합니다 — 기술, 재료 또는 원리

CC0 퍼블릭 도메인

이 블루프린트는 CC0로 공개되었습니다. 어떤 목적으로든 자유롭게 복사, 수정, 배포 및 사용할 수 있습니다.

제품 구매를 통해 메이커를 지원하세요. 판매자가 설정한 메이커 커미션 을 받거나, 이 블루프린트의 새로운 반복을 만들어 연결로 포함시킬 수 있습니다.

토론

(0)

로그인 하여 토론에 참여하세요

댓글 로딩 중...